Technical Specifications |
Test Range |
2.5 mm to 5 meter (in steel). In Hot Key mode it has 13 preset values. Fine mode adjustable in step of 1mm. |
Velocity |
1000 m/s to 9999 m/s. In Hot Key mode it has 7 preset values. Fine mode adjustable in step of 1 m/s. |
Delay |
Up to 3 meter adjustable in 0.5mm or 5mm step. |
Gain |
100 dB calibrated gain adjustable in 0.1, 0.5, 1, 2, 6 or 14 dB step. |
Rejection |
0 to 100% FSH with LED indicator. |
Rectification |
Fullwave rectified with filtering. |
Frequency |
0.5 MHz to 15 MHz (Wideband). |
Linearity deviation |
Vertical: 3%,Horizontal: 0.5%. |
Test Modes |
Pulse echo and transmit/receive. |
Connectors |
LEMO Size 00 (2 nos). |
Monitor |
Dual gate adjustable in 1% of Screen width with Positive/Negative logic, Gate Expand modes. |
Gate Expand |
Expands Range to width of the gate. |
A-Scan memory |
500 Trace Patterns can be stored, recalled or transferred to PC via USB port. A-Scan trace pattern can be directly stored on SD card or on External USB disk. |
Calibration Set-up |
50 different calibration set-ups can be Stored and Recalled. |
Software |
ArjnSoft Interface software for transferring A-Scan from Arjun10 to PC is supplied. |
USB and SD card |
External USB disk and Micro SD card (upto 8GB) ports provided for data storage. |
Display |
5 inch Colour Wide VGA LCD/TFT display (800×480 pixel, size 108×64 mm) with higher resolution LCD for better signal representation and better visibility with different colour sets. |
Reference A-Scan |
Reference A-Scan pattern of standard test object can saved and recalled in the background for easy comparisons during testing. |
Echo Dynamic |
Echo Dynamic pattern display. |
DAC |
Dynamic DAC curve can be Digitally plotted(Smooth parabolic curve) on screen with selectable additional offset curves from 0 to 14 dB in 0.5 dB selectable steps. DAC curve can be set as flaw monitor gate. DAC curve can be plotted using minimum 2 to maximum 10 points. |
Digital Readout And Trig. Function |
Thickness/Depth can be displayed in digital readout when using a normal probe and Beam path, Surface Distance and Depth are directly displayed when angle probe is in use. |
Update Rate |
50Hz. |
Power |
Lithium-Ion Battery pack 7.4VDC,5AH, gives 10 hours continuous operation from fully charged battery and External charger is provided to charge the battery with out engaging the main unit. |
Charger |
Input Voltage 100 to 240 VAC. Charge status indicator is provided. |
Temperature |
0 to 55°C. |
Size |
117 x 176 x 56 mm (HxWxD). |
Weight |
800 gms with Battery. |