Technical Specifications |
Test Range |
10mm to 1000mm (@5920 M/sec Steel Velocity equivalent) continuously variable. It can be adjusted with hot key for coarse range step. Horizontal linearity 1%. |
Velocity |
1000m/s to 9999m/s. continuous adjustment. It can be adjusted using hot key in 7 preset values. |
Delay |
0-1000mm Continuous variable. It can be adjusted using hot key in coarse delay step. |
Gain |
0-80 db with 0.1,0.5,1,2,6,12 dB step selection. Vertical Linearity: 5% |
Rejection |
Linear type (0-99%) of full screen height, 1% step. |
Rectification |
Full wave rectified A-Scan. |
Frequency |
0.2 to 20 MHz |
No of channels |
16 channels with separate A-Scan traces, Any or all can be selected. |
Test Modes |
Pulse echo or transmit/receive. |
Transmitter |
Negative Square wave type. |
Freeze/Peak Freeze |
Current display Freeze and Peak freeze to create echo dynamic pattern. |
Connector |
BNC type connectors for all channels. |
Monitor |
Dual gate for each channel adjustable in 1% screen width with Positive/Negative logic, Gate Expand modes. |
Gate Expand |
Expands Range to width of the gate. |
A-Scan memory (built-in) |
200 Multicolored A-Scan with calibration parameters with measured value. Using external USB pen drive virtually unlimited A-scan and Calibration set-ups can be stored. |
Calibration Set-up |
100 calibration parameters set-ups. |
Software |
Suitable PC software provided to review/ print report of the stored A-scan files. |
Display |
TFT Color with LED backlight 152mm x 91 mm Viewing area, 800 x 480 pixels. All channels have separate base line with color selectable by user. |
A-Scan Display |
Envelop or filled A-scan Pattern. |
DAC |
Separate Dynamic DAC curve for each channel with additional -6 and -14 dB. DAC curve can be set as flaw monitor gate. Using 2 to 10 points DAC curve can be plotted. |
Measurement |
Echo amplitude, Sound path, Echo to Echo distance. When angle probe is used then using trigonometric functions it calculates and displays Surface and Depth distance of defect. |
Measuring Unit |
Millimeters. |
PRF |
Maximum 8000 Hz which varies in step of 50 Hz as per parameter set value due to Auto limiting. |
Update Rate |
50Hz |
Power Source |
Operates on Built in Li-Ion rechargeable battery or on Mains using supplied charger. Operation time is 8 hours with fully charged battery. |
Charger |
Input Voltage 100 to 240 VAC. Charge status indicator is provided. |
Battery charge status |
Battery charge status by suitable LED indication on charger. |
Battery Status |
Battery status indicator with auto low battery shut down to protect battery. |
Encoder |
Positional encoder interface to indicate travel distance/position of probe. |
Keyboard |
User friendly with direct access to frequently used functions. |
Operating Temperature |
0 to 55⁰ C. (For operation) |
Dimensions |
247mm x 151mm x 70mm |
Weight |
Approx 2 Kgs. |